Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 863: CAL ADJ /5500 INSTRUMENT: Fluke 865: CAL ADJ /5500 INSTRUMENT: Fluke 867: CAL ADJ /5500 DATE: 21-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 320 CONFIGURATION: Fluke 5500A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # # Source: # Fluke 86X Calibration Procedure, Preliminary, February 1995 # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # # The following describes the use of the [I] special constructions # used in the PORT FSC in this procedure: # # SF (Set Function) command # # 6.003 PORT SF 9[13][I][D1000] # # [I] reads the acknowledge response # 1.001 ASK- W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP [32] WARNING 1.003 DISP HIGH VOLTAGE is used or exposed during the performance 1.003 DISP of this calibration. DEATH ON CONTACT may result if 1.003 DISP personnel fail to observe safety precautions. 1.004 DISP Plug external power adapter into UUT to insure stable 1.004 DISP power while calibration is being performed. 1.005 DISP If at any time, you wish to abort the calibration, 1.005 DISP select the MET/CAL Terminate option and "Abort CAL" 1.005 DISP (Soft Key 5) on the UUT. 1.006 DISP 10 minute warm-up time requirement. 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 90%. 1.007 HEAD STEP 1: mVDC CALIBRATION ADJUSTMENT 1.008 DISP Connect the 5500A to the UUT as follows: 1.008 DISP [32] 5500A NORMAL HI to UUT Volt Ohm Diode Input 1.008 DISP [32] 5500A NORMAL LO to UUT COM Input 1.009 DISP Rotate the UUT function switch to mVDC. 1.009 DISP 1.009 DISP Place the UUT in calibration mode by pressing the. 1.009 DISP recessed calibration button located below the power 1.009 DISP adapter plug on the right side of the UUT. 1.009 DISP 1.009 DISP Select Advance when the calibration display appears. 1.010 5500 900mV S 2W 1.011 DISP Press Proceed (Soft Key 1) when reading has settled. 1.012 DISP Select Advance when Step 2 appears on the left side of 1.012 DISP the UUT display. 1.013 HEAD STEP 2: mVDC CALIBRATION ADJUSTMENT 1.014 5500 2.9V S 2W 1.015 DISP Press Proceed (Soft Key 1) when reading has settled. 1.016 DISP Select Advance when Step 3 appears on the left side of 1.016 DISP the UUT display. 1.017 HEAD STEP 3: mVDC CALIBRATION ADJUSTMENT 1.018 5500 290mV S 2W 1.019 DISP Press Proceed (Soft Key 1) when reading has settled. 1.020 DISP Select Advance when Step 4 appears on the left side of 1.020 DISP the UUT display. 1.021 5500 * S 1.022 HEAD STEP 4: VDC CALIBRATION ADJUSTMENT 1.023 DISP Rotate the UUT function switch to VDC. 1.023 DISP 1.023 DISP Select Advance when the calibration display appears. 1.024 5500 1000V S 2W 1.025 DISP Press Proceed (Soft Key 1) when reading has settled. 1.026 DISP Select Advance when Step 5 appears on the left side of 1.026 DISP the UUT display. 1.027 HEAD STEP 5: VDC CALIBRATION ADJUSTMENT 1.028 5500 290V S 2W 1.029 DISP Press Proceed (Soft Key 1) when reading has settled. 1.030 DISP Select Advance when Step 6 appears on the left side of 1.030 DISP the UUT display. 1.031 HEAD STEP 6: VDC CALIBRATION ADJUSTMENT 1.032 5500 29V S 2W 1.033 DISP Press Proceed (Soft Key 1) when reading has settled. 1.034 DISP Select Advance when Step 7 appears on the left side of 1.034 DISP the UUT display. 1.035 5500 * S 1.036 HEAD STEP 7: mADC CALIBRATION ADJUSTMENT 1.037 DISP Connect the 5500A to the UUT as follows: 1.037 DISP [32] 5500A AUX HI to UUT mA/uA Input 1.037 DISP [32] 5500A AUX LO to UUT COM Input 1.038 DISP Rotate the UUT function switch to mA/uA. 1.038 DISP 1.038 DISP Select Advance when the calibration display appears. 1.039 5500 290mA S 2W 1.040 DISP Press Proceed (Soft Key 1) when reading has settled. 1.041 DISP Select Advance when Step 8 appears on the left side of 1.041 DISP the UUT display. 1.042 HEAD STEP 8: mADC CALIBRATION ADJUSTMENT 1.043 5500 29mA S 2W 1.044 DISP Press Proceed (Soft Key 1) when reading has settled. 1.045 DISP Select Advance when Step 9 appears on the left side of 1.045 DISP the UUT display. 1.046 HEAD STEP 9: mADC CALIBRATION ADJUSTMENT 1.047 5500 2.9mA S 2W 1.048 DISP Press Proceed (Soft Key 1) when reading has settled. 1.049 DISP Select Advance when Step 10 appears on the left side of 1.049 DISP the UUT display. 1.050 HEAD STEP 10: mADC CALIBRATION ADJUSTMENT 1.051 5500 290uA S 2W 1.052 DISP Press Proceed (Soft Key 1) when reading has settled. 1.053 DISP Select Advance when Step 11 appears on the left side of 1.053 DISP the UUT display. 1.054 HEAD STEP 11: mADC CALIBRATION ADJUSTMENT 1.055 5500 29uA S 2W 1.056 DISP Press Proceed (Soft Key 1) when reading has settled. 1.057 DISP Select Advance when Step 12 appears on the left side of 1.057 DISP the UUT display. 1.058 5500 * S 1.059 HEAD STEP 12: ADC CALIBRATION ADJUSTMENT 1.060 DISP Connect the 5500A to the UUT as follows: 1.060 DISP [32] 5500A AUX HI to A Input 1.060 DISP [32] 5500A AUX LO to UUT COM Input 1.061 DISP Rotate the UUT function switch to A. 1.061 DISP 1.061 DISP Select Advance when the calibration display appears. 1.062 5500 10A S 2W 1.063 DISP Press Proceed (Soft Key 1) when reading has settled. 1.064 DISP Select Advance when Step 13 appears on the left side of 1.064 DISP the UUT display. 1.065 HEAD STEP 13: ADC CALIBRATION ADJUSTMENT 1.066 5500 2.9A S 2W 1.067 DISP Press Proceed (Soft Key 1) when reading has settled. 1.068 DISP Select Advance when Step 14 appears on the left side of 1.068 DISP the UUT display. 1.069 HEAD STEP 14: ADC CALIBRATION ADJUSTMENT 1.070 5500 0.29A S 2W 1.071 DISP Press Proceed (Soft Key 1) when reading has settled. 1.072 DISP Select Advance when Step 15 appears on the left side of 1.072 DISP the UUT display. 1.073 5500 * S 1.074 HEAD STEP 15: Ohms/nS CALIBRATION ADJUSTMENT 1.075 DISP Connect the 5500A to the UUT as follows: 1.075 DISP [32] 5500A NORMAL HI to UUT Volt Ohm Diode Input 1.075 DISP [32] 5500A NORMAL LO to UUT COM Input 1.076 DISP Rotate the UUT function switch to Ohms. 1.076 DISP 1.076 DISP Select Advance when the calibration display appears. 1.077 5500 2.9MZ S 2W 1.078 DISP Press Proceed (Soft Key 1) when reading has settled. 1.079 DISP Select Advance when Step 16 appears on the left side of 1.079 DISP the UUT display. 1.080 HEAD STEP 16: Ohms/nS CALIBRATION ADJUSTMENT 1.081 5500 290kZ S 2W 1.082 DISP Press Proceed (Soft Key 1) when reading has settled. 1.083 DISP Select Advance when Step 17 appears on the left side of 1.083 DISP the UUT display. 1.084 5500 * S 1.085 HEAD STEP 17: Ohms/nS CALIBRATION ADJUSTMENT 1.086 DISP Connect the 5500A to the UUT as follows: 1.086 DISP [32] 5500A SENSE HI to UUT Volt Ohm Diode Input 1.086 DISP [32] 5500A SENSE LO to UUT COM Input 1.086 DISP [32] 5500A NORMAL HI to UUT Volt Ohm Diode Input 1.086 DISP [32] 5500A NORMAL LO to UUT COM Input 1.087 5500 29kZ S CW 1.088 DISP Press Proceed (Soft Key 1) when reading has settled. 1.089 DISP Select Advance when Step 18 appears on the left side of 1.089 DISP the UUT display. 1.090 HEAD STEP 18: Ohms/nS CALIBRATION ADJUSTMENT 1.091 5500 2.9kZ S CW 1.092 DISP Press Proceed (Soft Key 1) when reading has settled. 1.093 DISP Select Advance when Step 19 appears on the left side of 1.093 DISP the UUT display. 1.094 HEAD STEP 19: Ohms/nS CALIBRATION ADJUSTMENT 1.095 5500 290Z S CW 1.096 DISP Press Proceed (Soft Key 1) when reading has settled. 1.097 DISP Select Advance when Step 20 appears on the left side of 1.097 DISP the UUT display. 1.098 5500 * S 1.099 HEAD STEP 20: Ohms/nS CALIBRATION ADJUSTMENT 1.100 DISP Connect the 5500A to the UUT as follows: 1.100 DISP [32] 5500A NORMAL HI to UUT Volt Ohm Diode Input 1.100 DISP [32] 5500A NORMAL LO to UUT COM Input 1.101 5500 3.5MZ S 2W 1.102 DISP Press Proceed (Soft Key 1) when reading has settled. 1.103 DISP Select Advance when Step 21 appears on the left side of 1.103 DISP the UUT display. 1.104 HEAD STEP 21: Ohms/nS CALIBRATION ADJUSTMENT 1.105 5500 350kZ S 2W 1.106 DISP Press Proceed (Soft Key 1) when reading has settled. 1.107 DISP Select Advance when Step 22 appears on the left side of 1.107 DISP the UUT display. 1.108 5500 * S 1.109 HEAD STEP 22: VAC CALIBRATION ADJUSTMENT 1.110 DISP Rotate the UUT function switch to VAC. 1.110 DISP 1.110 DISP Select Advance when the calibration display appears. 1.111 5500 1000V 1kH SI S 2W 1.112 DISP Press Proceed (Soft Key 1) when reading has settled. 1.113 DISP Select Advance when Step 23 appears on the left side of 1.113 DISP the UUT display. 1.114 HEAD STEP 23: VAC CALIBRATION ADJUSTMENT 1.115 5500 100V 1kH SI S 2W 1.116 DISP Press Proceed (Soft Key 1) when reading has settled. 1.117 DISP Select Advance when Step 24 appears on the left side of 1.117 DISP the UUT display. 1.118 HEAD STEP 24: VAC CALIBRATION ADJUSTMENT 1.119 5500 290V 1kH SI S 2W 1.120 DISP Press Proceed (Soft Key 1) when reading has settled. 1.121 DISP Select Advance when Step 25 appears on the left side of 1.121 DISP the UUT display. 1.122 HEAD STEP 25: VAC CALIBRATION ADJUSTMENT 1.123 5500 29V 1kH SI S 2W 1.124 DISP Press Proceed (Soft Key 1) when reading has settled. 1.125 DISP Select Advance when Step 26 appears on the left side of 1.125 DISP the UUT display. 1.126 HEAD STEP 26: VAC CALIBRATION ADJUSTMENT 1.127 5500 2.9V 1kH SI S 2W 1.128 DISP Press Proceed (Soft Key 1) when reading has settled. 1.129 DISP Select Advance when Step 27 appears on the left side of 1.129 DISP the UUT display. 1.130 HEAD STEP 27: VAC CALIBRATION ADJUSTMENT 1.131 5500 290mV 1kH SI S 2W 1.132 DISP Press Proceed (Soft Key 1) when reading has settled. 1.133 DISP Select Advance when Step 28 appears on the left side of 1.133 DISP the UUT display. 1.134 HEAD STEP 28: VAC CALIBRATION ADJUSTMENT 1.135 5500 29mV 1kH SI S 2W 1.136 DISP Press Proceed (Soft Key 1) when reading has settled. 1.137 DISP Select Advance when Step 29 appears on the left side of 1.137 DISP the UUT display. 1.138 5500 * S 1.139 HEAD STEP 29: mAAC CALIBRATION ADJUSTMENT 1.140 DISP Connect the 5500A to the UUT as follows: 1.140 DISP [32] 5500A AUX HI to UUT mA/uA Input 1.140 DISP [32] 5500A AUX LO to UUT COM Input 1.141 DISP Rotate the UUT function switch to mA/uA. 1.141 DISP 1.141 DISP Select Advance when the calibration display appears. 1.142 5500 290mA 1kH SI S 2W 1.143 DISP Press Proceed (Soft Key 1) when reading has settled. 1.144 DISP Select Advance when Step 30 appears on the left side of 1.144 DISP the UUT display. 1.145 HEAD STEP 30: mAAC CALIBRATION ADJUSTMENT 1.146 5500 29mA 1kH SI S 2W 1.147 DISP Press Proceed (Soft Key 1) when reading has settled. 1.148 DISP Select Advance when Step 31 appears on the left side of 1.148 DISP the UUT display. 1.149 HEAD STEP 31: mAAC CALIBRATION ADJUSTMENT 1.150 5500 2.9mA 1kH SI S 2W 1.151 DISP Press Proceed (Soft Key 1) when reading has settled. 1.152 DISP Select Advance when Step 32 appears on the left side of 1.152 DISP the UUT display. 1.153 HEAD STEP 32: mAAC CALIBRATION ADJUSTMENT 1.154 5500 290uA 1kH SI S 2W 1.155 DISP Press Proceed (Soft Key 1) when reading has settled. 1.156 DISP Select Advance when Step 33 appears on the left side of 1.156 DISP the UUT display. 1.157 HEAD STEP 33: mAAC CALIBRATION ADJUSTMENT 1.158 5500 29uA 1kH SI S 2W 1.159 DISP Press Proceed (Soft Key 1) when reading has settled. 1.160 DISP Select Advance when Step 34 appears on the left side of 1.160 DISP the UUT display. 1.161 5500 * S 1.162 HEAD STEP 34: AC Amps CALIBRATION ADJUSTMENT 1.163 DISP Connect the 5500A to the UUT as follows: 1.163 DISP [32] 5500A AUX HI to A Input 1.163 DISP [32] 5500A AUX LO to UUT COM Input 1.164 DISP Rotate the UUT function switch to A. 1.164 DISP 1.164 DISP Select Advance when the calibration display appears. 1.165 5500 10A 1kH SI S 2W 1.166 DISP Press Proceed (Soft Key 1) when reading has settled. 1.167 DISP Select Advance when Step 35 appears on the left side of 1.167 DISP the UUT display. 1.168 HEAD STEP 35: AC Amps CALIBRATION ADJUSTMENT 1.169 5500 1A 1kH SI S 2W 1.170 DISP Press Proceed (Soft Key 1) when reading has settled. 1.171 DISP Select Advance when Step 36 appears on the left side of 1.171 DISP the UUT display. 1.172 HEAD STEP 36: AC Amps CALIBRATION ADJUSTMENT 1.173 5500 2.9A 1kH SI S 2W 1.174 DISP Press Proceed (Soft Key 1) when reading has settled. 1.175 DISP Select Advance when Step 37 appears on the left side of 1.175 DISP the UUT display. 1.176 HEAD STEP 37: AC Amps CALIBRATION ADJUSTMENT 1.177 5500 290mA 1kH SI S 2W 1.178 DISP Press Proceed (Soft Key 1) when reading has settled. 1.179 DISP Select Advance when Step 38 appears on the left side of 1.179 DISP the UUT display. 1.180 5500 * S 1.181 HEAD STEP 38: mVDC CALIBRATION ADJUSTMENT 1.182 DISP Connect the 5500A to the UUT as follows: 1.182 DISP [32] 5500A NORMAL HI to Volt Ohm Diode Input 1.182 DISP [32] 5500A NORMAL LO to UUT COM Input 1.183 DISP Rotate the UUT function switch to mVDC. 1.183 DISP 1.183 DISP Select Advance when the calibration display appears. 1.184 5500 29mV S 2W 1.185 DISP Press Proceed (Soft Key 1) when reading has settled. 1.186 DISP Select Save Cal. to store new constants or Abort Cal. 1.186 DISP when calibration completes. 1.187 END